4
PROGRAM INSTALLATION & CONTROL
Parr Instrument Company
24
Calorimeter Operations
Operating Mode Determination
Bomb Installed/EE 1/940.0
Heater and Pump OFF
Operating Controls
Method of Operation Dynamic
Reporting Units BTU/lb
Use Spiking Correction OFF
“OTHER” Multiplier 4.1868
Calibrate Touchscreen
LCD Backlight Timeout(s) 1200 S
LCD Backlight Intensity 70%
Print Error Messages ON
Language English
Spike Controls
Use Spiking OFF
Heat of Combustion of Spike 6318.4
Use Fixed Spike OFF
Weight of Fixed Spike 0.0
Prompt for Spike before Weight OFF
Bomb Rinse Tank Control
Report Rinse Tank Empty ON
Rinse Tank Capacity 150
# Rinses Left 150
Reset Rinse Tank Counter
Rinse Time 25
Rinse Flush Time 20
Clear Time 100
# of Rinse Cycles 3
Program Information and Control
Date & Time Settings
Volume Level Adjust 85%
Software and Hardware Info
Settings Protect OFF
User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart
User/Factory Settings
User Setup ID 63-1138
Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings
Compare settings with Factory Defaults
Calibration Data & Controls
Calibration Run Limit 10
EE Max Std Deviation 0.0
Heat of Combustion of Standard 6318.4
Bomb Service Interval 500
Use Bomb 1
Control Chart Parameters
Charted Value HOC Standard
Process Sigma 0.1
Temp Rise High Warning 8.5
Temp Rise Low Warning 5.1
Bomb 1 Through 4
EE Value 800.0
Protected EE Value OFF
Thermochemical Corrections Standardization
Fixed Fuse ON 50.0
Acid Correction Fixed HNO
3
8.0
Fixed Sulfur ON 0.0
Determination
Fixed Fuse ON 50.0
Acid Correction Fixed HNO
3
8.0
Fixed Sulfur OFF 0.0
Net Heat/Dry Factors
Calculation Factors
Nitric Acid Factor 1.58
Figure 4-1
6300 Factory Default Settings