650M6750Proximate InterfaceOperating Instruction ManualFor models produced after January 2012
Proximate Interfacewww.parrinst.com9Menu Operating Pages The 6750 Proximate Interface contains the fol-lowing keys that lead to sub-menus.Process Tim
Proximate InterfaceParr Instrument Company10Operating Controls (Continued)LCD Backlight Timeout: The unit is equipped with an automatic circuit to shu
Proximate Interfacewww.parrinst.com11User/Factory Settings:This key leads to a sub-menu that allows the user to save or recall user defi ned instrument
Proximate InterfaceParr Instrument Company12Data Entry Controls (Continued)Data Sources:This key accesses a sub-menu where the user can select input d
Proximate Interfacewww.parrinst.com13Communication Controls Menu Accesses sub-menus which set the communica-tions protocols for the printer and balan
Proximate InterfaceParr Instrument Company14Communication Controls Menu (Continued) » Balance Handler Strings. This key leads to a submenu that allows
Proximate Interfacewww.parrinst.com15 The following screen shot illustrates the contents of the Proximate Interface data directory as presented be a
Proximate InterfaceParr Instrument Company16 You can drag and drop or copy and paste test data fi les (with csv suffi x) from the web browser window to
Proximate Interfacewww.parrinst.com17 Where 10.1.5.32 is the IP address of the Proximate Interface. The following screen shot illustrates the Proxim
Proximate InterfaceParr Instrument Company18 From the Confi guration Page the Error Reporting and Sample ID & Bomb ID options can be set. Changes
Proximate Interfacewww.parrinst.com19 On the Run Data screen test data can be reviewed and fi ltered by status and type.
Proximate InterfaceParr Instrument Company20The System Info page displays the Software and Hardware Information page of the 6750.
Proximate Interfacewww.parrinst.com21File Management MenuRun Data File Manager:This key activates the File Manager. The File Man-ager is used to dele
Proximate InterfaceParr Instrument Company22DiagnosticsView System Log :This key is used to display the contents of /fl ash/log/messages. This fi le is
Proximate Interfacewww.parrinst.com23Basic Data Entry Procedures Pressing the SAMP ID key starts the data entry sequence. This will prompt the operat
Proximate InterfaceParr Instrument Company24Air Dried Loss or Total Moisture Entry Air Dried Loss (ADL) or Total Moisture (TM) weights are identifi ed
Proximate Interfacewww.parrinst.com25Batch Input Mode The block input mode consists of allowing the 6750 Proximate Interface to automatically prompt
Proximate InterfaceParr Instrument Company26Reporting Instructions The 6750 Proximate Interface can transmit for-matted data over the USB port to a 4
Proximate Interfacewww.parrinst.com27Memory Management The 6750 Calorimeter will hold data for 1000 tests in its memory. These tests may be pre weig
Proximate InterfaceParr Instrument Company28Data & Communication Specifi cations6750 Run Data Template The following list illustrates the order in
Proximate InterfaceParr Instrument Company2Table of ContentsPrefaceRelated Instructions — 3Scope — 3Concept of Operations — 3Customer Service — 3Safet
Proximate Interfacewww.parrinst.com29Filename Convention Test data fi les are named with the following con-vention. Test type FilenamePreliminary <
Proximate InterfaceParr Instrument Company30Sartorius Balance Interface (Continued) The interface will send this command string once after the PROCEE
Proximate Interfacewww.parrinst.com31References 1. ASTM D2013 Standard Method of preparing Coal Samples for Analysis.2. ASTM D3172 Standard Practice
Proximate InterfaceParr Instrument Company32FiguresFigure 16750 Proximate Interface - Internal Parts ViewW/(4) SA1140RP06351C2139E14 FUSECPU BOARDW/(4
Proximate Interfacewww.parrinst.com33Figure 26750 Proximate Interface - Back ViewFigure 36750 Proximate Cover Panel - Internal ViewDISPLAY(4) TN1140HL
Proximate InterfaceParr Instrument Company34Figure 46750 Proximate Interface SchematicJ10J8J1BACK PANELSPEAKERCABLE ASSYUSB PORT 1J6A2167E USB PERIPHE
Revision 05/18/12
Proximate Interfacewww.parrinst.com3Concept of Operations The 6750 Proximate Interface is an advanced microprocessor system designed to:t Collect an
Proximate InterfaceParr Instrument Company4Preface (Continued) The 6750 Proximate Interface does not require a special oven, furnace or analytical ba
Proximate Interfacewww.parrinst.com5Explanation of SymbolsI On positionO Off Position~ Alternating Current (AC)This CAUTION symbol may be present on t
Proximate InterfaceParr Instrument Company6Installation This apparatus is to be used indoors. It requires at least 2 square feet of workspace on a st
Proximate Interfacewww.parrinst.com7Menu Keys The controls that change the data fi eld informa-tion in the menus will be one of the following:tToggle
Proximate InterfaceParr Instrument Company8Factory Default SettingsProcess TimersMoisture 01:00:00Ash 04:00:00VM 00:07:00Operating ControlsDevice Type