6725 Semi-micro Calorimeter Instruction Manual
14-1
TABLES
Table 1 – Default Settings
Note: The following are recommended
operation settings for the 6725 Semi-micro
Calorimeter.
Calorimeter Operation
Operating Mode Determination
Bomb Installed/EE 1/2400
Stirrer OFF
Operating Controls
Method of Operation Combustion
Reporting Units Btu/lb
Use Spiking Correction OFF
“OTHER” Multiplier 4.1868
LCD Backlight Timeout(s) 1200 S
LCD Backlight Intensity 70%
Print Error Messages ON
Language English
Spiking Correction
Use Spiking OFF
Heat of Combustion of Spike 6318.4
Use Fixed Spike OFF
Weight of Fixed Spike 0.0
Prompt for Spike before Weight OFF
Jacket Temp. Source
Jacket Temp. Source Calculated
Fixed Jacket Temp 25.0
Program Info and Control
Settings Protect OFF
Calibration Data & Controls
Calibration Run Limit 10
EE Max Std Deviation 0.0
Heat of Combustion
of Standard 6318.4
Bomb Service Interval 500
Use Bomb 1
Bomb 1 Through 4
EE Value 2400
Protected EE Value OFF
Thermochemical Corrections
Standardization
Fixed Fuse Correction ON 15
Acid Correction Fixed HNO
3
10.0
Fixed Sulfur Correction ON 0.0
Determination
Fixed Fuse Correction ON 15
Acid Correction Fixed HNO
3
10.0
Fixed Sulfur Correction OFF 0.0
Calculate Net Heat of Combustion OFF
Calculation Factors
Nitric Acid Factor 1.58
Acid Multiplier 0.0709
Sulfur Value is Percent ON
Sulfur Multiplier 0.6238
Fuse Multiplier 1.0
Use Offset Correction (ISO) OFF
Offset Value 0.0
Net Heat/Dry Factors
Net Heat & Dry Disable
Fixed Hydrogen OFF 0.0
Fixed Oxygen ON 0.0
Fixed Nitrogen ON 0.0
Calculate Net Heat of Combustion OFF
Fixed Moisture as Determined OFF 0.0
Fixed Moisture as Received OFF 0.0
Dry Calculation OFF
Data Entry Controls
Prompt for Bomb ID ON
Weight Entry Mode Touch Screen
Acid Entry Mode Touch Screen
Auto Sample ID Controls ON
Sample Weight Warning above 2.0
Spike Weight Entry Mode Touch Screen
Sulfur Entry Mode Touch Screen
Moisture Entry Mode Touch Screen
Auto Preweigh Controls ON
Auto Sample ID Controls
Automatic Sample ID ON
Automatic Sample ID Number 1
Automatic Sample ID Increment 1