6200Calorimeter O
p
eratin
g
Instruction Manual
Sartorious
Balance
Interface
Field Length
polarity 1
space 1
data 8
space 1
stability 2
CR 1
LF 1
The polarity field must contain either a
“+” or a space. Leading zeros in the data
field are blanked, except for the one to the
left of the decimal point. The stability field
must contain “g_” for the calorimeter to
accept a mass. The balance should be
configured to transmit data upon receipt of
the following command string:
[ESC] P [CR] [LF]
Note:
The automatic data output option
should not be used.
The calorimeter will send this command
string once every few seconds after the
ENTER key has been pressed during a
mass entry sequence. The ENTER key
should only be pressed when the mass
reading is stable. However, unstable
readings will be rejected and a warning will
be issued. Acknowledging the warning will
re-issue the command string to the balance
on a periodic basis.
D-2
APPENDIX D