Parr-instrument 6100 Manual do Utilizador Página 1

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Parr-instrument 6100. Parr Instrument 6100 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Operating Instruction Manual

6100Compensated Jacket CalorimeterOperating Instruction ManualFor models produced after October 2010584M

Página 2

2InstallationParr Instrument Company8chaPter 2InstallationEnvironmental ConditionsThe 6100 Calorimeter is completely assembled and given a thorough te

Página 3 - Table of Contents6100

Installation61002www.parrinst.com9procedure and calculations. The user should config-ure the calorimeter at this time to accommodate the desired sample

Página 4 - Table of Contents

2InstallationParr Instrument Company10Note: The Cooling Water Outlet, Cooling Water Inlet, Tank Fill and Tank Drain ports are not used on the 6100 Cal

Página 5 - Preface

www.parrinst.com11Notes 61002

Página 6 - General Specifications

3Instrument DescriptionParr Instrument Company12chaPter 3Instrument DescriptionTypes of ControlsAll calorimeter configurations and operations are handl

Página 7 - Getting Started

Instrument Description61003www.parrinst.com13Menu KeysThe controls that change the data field information in the menus will be one of the following:1.

Página 8 - Concept of Operation

4Program Installation & ControlParr Instrument Company14chaPter 4Program Installation & ControlSoftware InstallationThe program in the 6100 Ca

Página 9

www.parrinst.com15Notes 61004

Página 10 - Installation

4Program Installation & ControlParr Instrument Company16Calorimeter OperationsOperating Mode DeterminationBomb Installed/EE 1/2400.0Operating Cont

Página 11

Program Installation & Control61004www.parrinst.com17Net Heat/Dry FactorsFixed Hydrogen OFF 0.0Fixed Oxygen ON 0.0Fixed Nitrogen ON 0.0Calculate N

Página 13 - Notes

5Operating InstructionsParr Instrument Company18chaPter 5Operating InstructionsOperating the 1108P Oxygen Combustion VesselDetailed instructions for p

Página 14 - Instrument Description

Operating Instructions61005www.parrinst.com194. Set the bucket in the calorimeter. Attach the lifting handle to the two holes in the side of the scre

Página 15

5Operating InstructionsParr Instrument Company20Samples and Sample HoldersParticle Size and Moisture Content. Solid samples burn best in an oxygen bom

Página 16 - Revising Default Settings

Operating Instructions61005www.parrinst.com21Capsules should be monitored for wear. Do not use the capsule if the wall or base thickness is less than

Página 17

5Operating InstructionsParr Instrument Company22Explosives and High Energy FuelsThe 1108P and 1108PCL vessels used in the 6100 Calorimeter have been d

Página 18 - Table 4-1

Operating Instructions61005www.parrinst.com23Volatile samples are defined as one with an initial boiling point below 180 ºC per ASTM D-2.Low volatile s

Página 19

6Corrections & Final ReportsParr Instrument Company24chaPter 6Corrections & Final ReportsEntering Corrections and Obtaining the Final Report F

Página 20 - Operating Instructions

www.parrinst.com25Notes 61006

Página 21

7Reporting InstructionsParr Instrument Company26chaPter 7Reporting InstructionsReport Option SectionThe 6100 Calorimeter can transmit its stored test

Página 22

Reporting Instructions61007www.parrinst.com27The displayed files can be sorted by sample ID num-ber, by type, by status or by date of test by simply to

Página 23

www.parrinst.comTable of Contents61001Preface 3Scope 3Related Instructions 3Explanation of Symbols 4Safety Information 4Intended Usage 4G

Página 24

8File ManagementParr Instrument Company28chaPter 8File ManagementThe 6100 Calorimeter will hold data for 1000 tests in its memory. These tests may be

Página 25

www.parrinst.com29Notes 61008

Página 26 - Fixed Corrections

9Maintenance & TroubleshootingParr Instrument Company30chaPter 9Maintenance & TroubleshootingOxygen BombUnder normal usage the 1108P Parr Oxy

Página 27

Maintenance & Troubleshooting61009www.parrinst.com316100 Calorimeter Error ListThe calorimeter will run a number of diagnostic checks upon itself

Página 28 - Reporting Instructions

AMenu Operating InstructionsParr Instrument Company32aPPendix a Menu Operating Instructions When the START key is pressed, the calorimeter will promp

Página 29

Menu Operating Instructions6100Awww.parrinst.com33Temperature Plot Setup MenuEnable Bucket: Toggles ON/OFF.Bucket Autoscale: Toggles ON/OFF.Enable J

Página 30 - File Management

AMenu Operating InstructionsParr Instrument Company34Operating Controls MenuMethod of Operation: Offers an operating mode of either dynamic or equilib

Página 31

Menu Operating Instructions6100Awww.parrinst.com35Program Information and Control MenuDate & Time: Accesses a sub-menu to set the cur-rent date

Página 32 - Oxygen Bomb

AMenu Operating InstructionsParr Instrument Company36Bomb Service Interval: Displays the maximum number of times a bomb may be fired before it is flagg

Página 33 - 6100 Calorimeter Error List

Menu Operating Instructions6100Awww.parrinst.com37Note: For rapid turn around between tests, user may wish to use two bombs. Each bomb should be assig

Página 34 - Menu Operating Instructions

Parr Instrument CompanyTable of Contents2Chapter 9 30Maintenance & Troubleshooting 30Oxygen Bomb 30Fuses 306100 Calorimeter Error List 3

Página 35

AMenu Operating InstructionsParr Instrument Company38Thermochemical Calculations MenuStandardization CorrectionFixed Fuse Correction: Displays both t

Página 36

Menu Operating Instructions6100Awww.parrinst.com39Net Heat/Dry Factors: Accesses the Net Heat/Dry Factors sub-menu, which provides for setting the ne

Página 37

AMenu Operating InstructionsParr Instrument Company40Data Entry Controls MenuPrompt for Bomb ID: In the ON position the control-ler will prompt for a

Página 38

Menu Operating Instructions6100Awww.parrinst.com41Sample Weight – Warning Above: This key displays and leads to a sub-menu used to set the maximum a

Página 39

AMenu Operating InstructionsParr Instrument Company42Reporting Controls MenuReport Width: The column width of the printer being used can be set to 40

Página 40

Menu Operating Instructions6100Awww.parrinst.com43Balance Port CommunicationsBalance Type: Toggles through the available balance templates. Balance P

Página 41

AMenu Operating InstructionsParr Instrument Company44File ManagementRun Data File Manager: This key activates the File Manager. The File Manager is

Página 42

www.parrinst.com45Notes 6100A

Página 43

BCalculationsParr Instrument Company46aPPendix BCalculationsCalculating the Heat of CombustionThe 6100 Calorimeter will automatically make all of the

Página 44

Calculations6100Bwww.parrinst.com47ASTM and ISO Methods DifferCurrent ASTM, ISO, and British Standard Methods differ on their treatment of the nitric

Página 45

Preface6100www.parrinst.com3Preface Tables Factory Default Settings 16Settings for ISO & BSI Methods 49Calorimeter Control Li

Página 46

BCalculationsParr Instrument Company48Acid and Sulfur Corrections• Total acid is the amount of base required to titrate the bomb washings (milliliter

Página 47

Calculations6100Bwww.parrinst.com49When the Acid Correction is set to Fixed Total the value is considered a final value and the operator is not prompte

Página 48 - Calculations

BCalculationsParr Instrument Company50ISO CalculationsBoth the ISO 1928 and BSI 1016: Part 5 methods for testing the calorific value of coal and coke,

Página 49

Calculations6100Bwww.parrinst.com51When using a spike the total weight of the spike and sample combined should not exceed 1.2 gram.Example 1: A cust

Página 50

CStandardizationParr Instrument Company52aPPendix cStandardizationStandardizing the CalorimeterThe Energy Equivalent FactorThe term “standardization”,

Página 51

Standardization6100Cwww.parrinst.com53Table C-1Calorimeter Control Limit Values in J/g When Benzoic Acid is Used as a Test SampleAccepted heat of comb

Página 52

CStandardizationParr Instrument Company54Table C-2Calorimeter Control Limit Values in cal/g When Benzoic Acid is Used as a Test SampleAccepted heat of

Página 53

Standardization6100Cwww.parrinst.com55Table C-3Calorimeter Control Limit Values in BTU/lb When Benzoic Acid is Used as a Test SampleAccepted heat of c

Página 54 - Standardization

DCommunications InterfacesParr Instrument Company56aPPendix dCommunications Interfaces USB Port for ConnectionThe 6100 Calorimeter is also equipped wi

Página 55

Communications Interfaces6100Dwww.parrinst.com57Generic InterfaceThe data field should consist of 9 numeric characters (0 through 9, +, - and space) te

Página 56

PrefaceParr Instrument Company4Explanation of SymbolsI On PositionO Off Position~ Alternating CurrentThis CAUTION symbol may be present on the Product

Página 57

DCommunications InterfacesParr Instrument Company58Network InterfaceCalorimeter test data can be transferred to an Eth-ernet network connected compute

Página 58 - Communications Interfaces

Communications Interfaces6100Dwww.parrinst.com59Samba Server Feature (Optional)Samba was originally developed as an implementa-tion of the SMB (Server

Página 59

DCommunications InterfacesParr Instrument Company60The following screenshot illustrates the contents of the calorimeter data directory as presented by

Página 60

Communications Interfaces6100Dwww.parrinst.com61The calorimeter offers a web server service. Test reports can be viewed with a web browser using a URL

Página 61

DCommunications InterfacesParr Instrument Company62Clicking on the Config button will display the screen below. Changes made on this screen will change

Página 62

Communications Interfaces6100Dwww.parrinst.com63Clicking on the Run Data button displays a list of reports currently in the instrument memory.Clicking

Página 63

DCommunications InterfacesParr Instrument Company64Clicking on the System Info button will display the screen below.

Página 64

Communications Interfaces6100Dwww.parrinst.com65Clicking on the LCD Snap Shot button will display the current menu screen displayed by the calorimeter

Página 65

DCommunications InterfacesParr Instrument Company66Clicking on the Documentation button will display the screen below. Clicking on any of the links wi

Página 66

Communications Interfaces6100Dwww.parrinst.com67Bar Code PortThe use of barcodes in the laboratory has become a highly accurate, rapid and inexpensive

Página 67

Preface6100www.parrinst.com5Environmental ConditionsOperating: 15 ºC to 30 ºC; maximum relative humid-ity of 80% non-condensing. Installation Category

Página 68

ETechnical ServiceParr Instrument Company68aPPendix eTechnical Service Contact Technical ServiceShould you need assistance in the operation or servic

Página 69

www.parrinst.com69Notes 6100E

Página 70 - Technical Service

FParts Lists & DrawingsParr Instrument Company70aPPendix fParts Lists & DrawingsPrincipal Assemblies in CalorimeterItem Description1108 Oxygen

Página 71

Parts Lists & Drawings6100Fwww.parrinst.com71Spare and Installation Parts ListItem Description20VB Valve Seat230A O-ring, Bomb Head, 2-3/8 ID238A

Página 72 - Parts Lists & Drawings

FParts Lists & DrawingsParr Instrument Company72Figure F-16100 Compensated Jacket Calorimeter Cutaway Front

Página 73

Parts Lists & Drawings6100Fwww.parrinst.com73Figure F-26100 Compensated Jacket Calorimeter Cutaway Rear

Página 74

FParts Lists & DrawingsParr Instrument Company74Figure F-3A1279DD2 Control Schematic J10P6P8USB PORTJ8J1CONNECTS IN UPPER PORTBACK PANELUSB PORTPI

Página 75

Parts Lists & Drawings6100Fwww.parrinst.com75Figure F-4A1278DD Oxygen Solenoid Assembly

Página 76

FParts Lists & DrawingsParr Instrument Company76Figure F-5A1284DD2 Stirrer Hub Assembly

Página 77

Parts Lists & Drawings6100Fwww.parrinst.com77Figure F-6Stirrer Motor Assembly

Página 78

1Concept of OperationParr Instrument Company6chaPter 1 Concept of Operation OverviewThe 6100 Calorimeter has been designed to provide the user with:•

Página 79

Parr Instrument Company78 Notes

Página 81

584M R02 12/09/13

Página 82 - 584M R02 12/09/13

Concept of Operation61001www.parrinst.com7an automatic rolling average calculation to support calibration maintenance and verification. Following the a

Comentários a estes Manuais

Sem comentários